Laboratory
IX: 2 weeks
Human
Cardiovascular Activity
Goal
·
To introduce
techniques for measuring
·
Heart sounds
·
Blood
pressure
·
Respiratory-rate
and volume
·
Human ECG
responses
·
3-lead
measurements
·
12-lead
measurements
·
Computation
of axis
·
To examine
body-surface potentials and contours
·
To examine
cardiovascular modulation
·
Dive
·
Aerobic
Exercise
·
Mental
activity
I. Heart sound
measurements
Auscultation of heart sounds means
to listen to and study the sounds arising from the heart as it pumps blood.
These are the result of vibrations caused by the opening and closing of the
valves in the heart, and by the blood rebounding against the ventricular and
blood vessel walls. Sounds may be heard by using a stethoscope, or monitored
more accurately using a phonocardiogram.
Using a stethoscope, listen to the heart of
a group member, paying special attention to the four major auscultatory areas
on the chest (shown below). Also record the heart sounds using the
piezoelectric transducer.

II. Blood
pressure measurement
The cardiovascular system consists of a
system of blood vessels, and the heart, which pumps the blood throughout the
system. Blood pressure is a common
physiological measurement performed on the cardiovascular system, and gives a
measurement of the pressure in the blood vessels during the cardiac cycle. Systolic pressure is the maximum pressure in
the arteries, and occurs when the blood is ejected from the ventricles into the
arteries. Diastolic pressure is the
minimum pressure in the arteries, and occurs when the heart is filling with
blood, and results from the recoil of the elasticity of the blood vessels. The normal range for systolic pressure for a
resting adult is 100-139mmHg, and the normal range for diastolic pressure for a
resting adult is 60-89mmHg. In this
experiment, you will
be measuring blood pressure by the auscultatory method which involves the use
of a stethoscope or microphone and a sphygmomanometer. You will observe the
sounds detected during blood pressure measurement, referred to as Kortokoff
Sounds. The pressure cuff is inflated
so that the sound of blood flowing through the arteries is no longer heard, which
means that it is at a higher pressure than the pressure in the artery. It is slowly allowed to deflate. The first sound occurs when the systolic
pressure is reached, and blood starts to flow through the artery again. When the sounds become muffled, the
diastolic pressure has been reached.
Two other
commonly measured blood pressure parameters are: 1) Pulse pressure - the difference between the systolic and
diastolic pressures. The normal value is 40 mm Hg and 2) Mean blood pressure
– the diastolic pressure plus one third of the pulse pressure. This is the
average effective pressure forcing blood through the circulatory system. The
normal value is 96 to 100 mm Hg. The mean blood pressure is a function of two
factors - cardiac output (CO) and total peripheral resistance (TPR). Peripheral
resistance depends on the calibre (diameter) of the blood vessels and the
viscosity of the blood.
Mean
BP = Cardiac output (ml/sec) x TPR
Cardiac
output (ml/min) = Heart rate/min x Stroke volume (ml)
Thus,
the measurement of blood pressure provides us with information on the heart's
pumping efficiency and the condition of the systemic blood vessels. In general,
we say that the systolic blood pressure indicates the force of contraction of
the heart, whereas the diastolic blood pressure indicates the condition of the
systemic blood vessels (for instance, an increase in the diastolic blood
pressure indicates a decrease in vessel elasticity).
Experimental procedure
After making measurements by hearing the
Kortokoff sounds, also use the piezoelectric pressure transducer to record the
sounds as you deflate the cuff. Can you
record the sounds in the oscilloscope. Finally, lift your arm by a specific
distance of say 1.5 ft (450 mm) and measure blood pressure. What would have
predicted the systolic pressure to be? How does this compare to the actual
results? For your predictions, assume that the specific gravity of blood is 1 and
that of mercury is 13.5. Knowing the
millimeter distance your head is above your heart, can you calculate what the
systolic pressure is in the arteries of your brain? Can you explain why the
increased gravitational force or “gs” during flight might cause a pilot to pass
out? Optional question: A 16’ giraffe typically has its heart some eight and
one-half feet above ground level. If the brain of a giraffe requires a systolic
pressure of 120 mm. Hg, what systolic pressure must be produced by the heart to
satisfy this demand?
III. Arterial
pulse wave
The blood pressure within an
artery varies during each cardiac cycle. The highest pressure (systolic)
occurs when the heart is in its relaxation phase and no blood is flowing through
the semilunar valves. The difference between the systolic and diastolic
pressures is called the pulse pressure. A recording of these changes is
called an arterial pulse wave. A normal pulse wave over the aorta is
shown in Figure 3. The dicrotic notch results when the aortic semilunar
valves close, causing the blood in the aorta to rebound against the arterial
walls to produce a slight elevation in pressure.
The magnitude and contour of the
arterial pulse wave are directly related to the stroke volume and inversely
related to the compliance (elasticity) of the arterial vessels. As the vessels
lose their compliance (as with age or in arteriosclerosis), the stroke volume
increases and the height of the pulse wave increases (pulse pressure
increases). An examination of the pulse wave can give valuable clues to the
functioning of the arteries and heart, as is seen in the abnormal waves
pictured in Figure 4.
The velocity of the pulse wave as
it travels down the artery is also an important clinical measurement. The
arterial pulse wave moves over the large arteries at a rate of 3 to 5 m/sec and
over the small arteries at 14 to 15 m/sec. The difference in velocity is
related to the compliance of the vessels - the less compliance a vessel has,
the faster the pulse wave will move over it (as in the small arteries). Thus, a
measurement of the velocity of the pulse wave can also provide useful
information about changes in the vessel's elasticity (compliance). The velocity
will vary with the age of the individual (table1).
Recording the Peripheral Pulse
In this experiment, you will
record the pulse wave from the tip of the finger; a peripheral pulse. It is
recorded using a photoelectric pulse transducer, which measures changes in
blood volume (plethysmography). A light source in the transducer
transilluminates the finger tip, and the photoconductor detects changes in
light intensity within the finger caused by pulsatile variations in blood
volume.
Experimental Procedures
1. With the subject seated, attach
the transducer snugly to the palmar surface of the middle finger. Record the
pulse for 20 seconds with the subject's arm resting on the table. Now have the
subject raise the transducer above his head (arm extended) for 30 seconds and
record the pulse during the last 10 seconds.
Then have the subject lower the
transducer (arm hanging at his side) for 30 seconds. Note the characteristics
of the pulse wave profile.
IV.
Electrocardiogram
Every
living cardiac cell undergoes a regular sequence of electrical changes that
initiate the contractile activity (systole) and the relaxation (diastole)
of the cell. Thus, the contraction of the heart is associated with a compound
action potential that is initiated at the sinus node and sweeps over the
conduction path of the heart, preceding the mechanical contraction of the
cardiac fibers. During this depolarization and repolarization of the
myocardium, a potential difference is created between different regions on the
surface of the heart. A separation of charge or potential difference is called
a dipole. The electrical potential of the dipole is conducted through an
electrolyte solution, such as the interstitial fluid and blood plasma, and
eventually reaches the surface of the skin. By placing electrodes on the skin
surface, we are able to detect and record the electrical activity over the
heart surface prior to its contraction. By measuring the potential changes in
various directions across the heart, it is possible to detect abnormalities.
Here is a diagram of normal heart
conduction.
The electrocardiogram
(EKG) is a graphic record of the action potentials of the heart. It is
recorded with an electrocardiograph, and the study of this cardiac
electrical activity is called electrocardiography.
Lead I. Right arm to left arm.
The right arm is connected to the negative
terminal of the electrocardiograph, and the left arm to the positive terminal.
When the right arm is negative to the left arm the record shows an upward
deflection. Thus, lead I measures the potential difference between the
electrodes on the left and right arms, or across the base of the heart. We can
use a lead switching box to change recording modes.
Lead II. Right arm to left leg.
The right arm is connected to the negative
terminal, and the left leg to the positive terminal. Thus, lead II measures the
potential difference between the left leg and the right arm, or along the long
axis of the heart from base to apex.
Lead III. Left arm to left leg.
The left arm is connected to the negative
terminal, and the left leg to the positive terminal. This combination allows
lead III to measure the potential difference between the left leg and left arm,
or along the left side of the heart.
The
sinoatrial (SA) node initiates the cardiac impulse (epicardium in this area
becomes negative first), and this wave of negativity sweeps over the heart.
Because the SA node is nearer the right arm, this arm becomes negative while
the left arm and left leg are still positive, and the deflection of the record
is upward in those leads (I and II). The left arm is closer to the SA node, so
in lead III the first deflection is also upward as the left arm becomes
negative in reference to the left leg.
Other
lead configurations that can be derived from the standard leads are the Augmented
unipolar limb leads (frontal plane):
Lead aVR: RA (+) to [LA & LF] (-) (Rightward)
Lead aVL: LA (+) to [RA & LF] (-) (Leftward)
Lead aVF: LF (+) to [RA & LA] (-) (Inferior)
In some experiments, unipolar (+) chest leads (horizontal plane) are also used as shown in the figure below. These leads are : V1: right 4th intercostal space, V2: left 4th intercostal space, V3: halfway between V2 and V4, V4: left 5th intercostal space, mid-clavicular line, V5: horizontal to V4, anterior axillary line and V6: horizontal to V5, mid-axillary line. In addition to these intercoastal leads, we can also measure the potentials on various points on the surface of the chest to obtain a “body surface potential map”.
In all
these experiments, we will use the “Wilson’s central terminal” as a reference
electrde. To obtain a Wilson’s central terminal, connect a 5K resistor in
series with the two arm leads. Connect the two ends of the resistor together
and use this as your reference electrode.
Lead placement diagrams






An example of a body surface map.

Components of Normal ECG
Complex
The normal ECG comprises of :
·
P wave. Represents the spread of electrical activity
(wave of negativity) over the atria after the initial depolarization of the SA
node.
Experimental procedures for ECG
recordings
1. Make baseline measurements of
the:
Does
the cycle length ever vary (arrhythmia)? Is there a change in the cycle length
(heart rate) with inspiration or expiration? Are any of the waves abnormal?
A PR interval (adult) greater the 0.2
sec is abnormal and indicates first degree heart block. In second degree heart
block, there are P waves that are not followed by QRS waves; this may occur
regularly or irregularly. Third degree heart block is a complete AV
dissociation in which P waves occur quite regularly but have no relation to R
waves. The normal duration of QRS complex is 0.08 to 0.12 sec. A duration of
more than this indicates bundle branch block, or that the beat has arisen in
one of the ventricles- a so called ventricular beat or extra systole. Variations in the T wave are quite numerous
and require an expert diagnosis. Inversions of the T wave are not abnormal,
especially in lead III. Elevation of the ST segment by more than 2mm is
associated with acute injury or anoxia
2. Measure the QRS
axis (How
to )
3. Obtain a
body-surface contour map of the peak deflection of the QRS complex. For these
measurements, use two-channel acquisition where one channel is always fixed in
position. Use the peak of the QRS of that channel to time-align the QRS peak of
the second channel. Now change the position of the second channel to different
positions along the front of the body and for each position, determine the
magnitude of the second channel, time aligned to the QRS peak of the first
channel. From this data, make a contour plot of the body-surface potential.
Hints for minimizing ECG
measurement errors:
V. Respiration
and Airflow measurements
To measure
respiratory-rate and respiratory volume, Connect the differential pressure
transducer MPX10DP using the pin-out diagram below. Hook the transducer up to a
mouth-piece and breathe into it. The output of this transducer measures air-flow.
Integrate the air-flow measurement to obtain lung volume. Use a calibration of 3.5 mv/Kpa assuming a
3 V power-supply. Measure tidal volume and inspiratory and expiratory reserve
volumes (see figure below).

V. Integrated
cardiovascular measurements
Hook up transducers to
simultaneously measure heart-rate (using arterial pulse), ECG (using any
one-lead) and respiratory-rate.
VI. Modulation of
Cardiovascular activity
1. Dive reflex
Dip your head in
ice-cold water and observe changes in your cardiovascular response. Record and
discuss your observations.
2. Aerobic exercise
Examine the effects of exercise on circulatory and respiratory
physiology. Obtain a baseline Measure blood pressure, ECG, heart rate,
Breathing rate for 30 seconds. Start exercise: 200-300 jumping jacks. Measure
ECG, heart rate, Blood pressure, Breathing rate, Measure for 30 seconds
immediately following exercise. Exercise should be only at a moderately
strenuous level unless you are an athelete who is comfortable with strenuous
exercise in which case you can exercise more. Measure after the start of
exercise. Exercise should continue for about 3 to 5 minutes or until you feel
that you are tired. Then measure the same quantities every minute. How long
does it take for the various measurements to return to baseline?
3. Mental activity
Pick a mental activity
that you believe will modulate cardiovascular activity. Formulate a hypothesis
about cardiovascular modulation and demonstrate evidence in favor of your
hypothesis.